Index |
WMGBS24VLF |
WMGBS36VLF |
WMGBS44VLF |
WMPAS44VLF |
WMGBS57VLF |
Power supply |
230 V (±10%)
10 A, 50/60 Hz |
230 V (±10%)
10 A, 50/60 Hz |
230 V (±10%)
10 A, 50/60 Hz |
110 V (100 V...127 V)
15 A, 50/60 Hz |
230 V (±10%)
10 A, 50/60 Hz |
Output voltage |
0…24 kVRMS VLF 0,1 Hz
(option: 0,05 Hz + 0,02 Hz)
± 0…34 kV DC |
0…36 kVRMS VLF 0,1 Hz
(option: 0,05 Hz + 0,02 Hz)
± 0…52 kV DC |
0…44 kVRMS VLF 0,1 Hz
(option: 0,05 Hz + 0,02 Hz)
± 0…62 kV DC |
0…44 kVRMS VLF 0,1 Hz
(option: 0,05 Hz + 0,02 Hz)
± 0…62 kV DC |
0…57 kVRMS VLF 0,1 Hz
(option: 0,05 Hz + 0,02 Hz)
± 0…62 kV DC |
Voltage
waveshape VLF |
similar sine-wave, symmetrical, with True RMS measurement |
Voltage
waveshape DC |
direct voltage, negative and positive polarity |
Overcurrent trip (DC) |
10 mA |
Max. testable cable length, max. capacitance (VLF)
*at a cable capacitance of approx. 0.25 μF/km
|
up to 60 km
(15 μF at 24 kVRMS,
0,02 Hz)* |
up to 60 km
(15 μF at 18 kVRMS,
0,02 Hz)* |
up to 60 km
(15 μF at 18 kVRMS,
0,02 Hz)* |
up to 60 km
(15 μF at 6 kVRMS,
0,02 Hz)* |
up to 60 km
(15 μF at 18 kVRMS,
0,02 Hz)* |
Max. load at max.
output voltage (VLF) and 0.1 Hz |
5 μF at 24 kVRMS |
2,4 μF at 36 kVRMS |
1,6 μF at 44 kVRMS |
1,0 μF at 44 kVRMS |
0,55 μF at 57 kVRMS |
Discharge - integrated
automatic discharge
device |
max. 9000 J |
max. 12500 J |
max. 12500 J |
max. 12500 J |
max. 12500 J |
Voltage measuring range |
-40…0…40 kV
accuracy ±1% |
-60…0…60 kV
accuracy ±1% |
-70…0…70 kV
accuracy ±1% |
-70…0…70 kV
accuracy ±1% |
-70…0…70 kV
accuracy ±1% |
Current measuring
ranges |
±0…100 μA / 1 mA / 10 mA |
Operating temperature |
-20…+45°C |
Storage temperature |
-25…+70°C |
Duty |
continuous operation |
PC interface |
USB stick |
Construction |
in two parts: operation unit and high voltage unit |
Dimensions
and weight
Operation unit |
37 x 34 x 20 cm
17 kg |
Dimensions
and weight
High voltage
unit |
40 x 41 x 24 cm
38 kg |
40 x 44 x 24 cm
48 kg |
40 x 44 x 24 cm
49 kg |
40 x 44 x 24 cm
49 kg |
40 x 44 x 24 cm